Abstract
Background: To compare results from different corneal astigmatism measurement instruments; to reconstruct corneal astigmatism from the postimplantation spectacle refraction and toric intraocular lens (IOL) power; and to derive models for mapping measured corneal astigmatism to reconstructed corneal astigmatism.
Methods: Retrospective single centre study involving 150 eyes treated with a toric IOL (Alcon SN6AT, DFT or TFNT). Measurements included IOLMaster 700 keratometry (IOLMK) and total keratometry (IOLMTK), Pentacam keratometry (PK) and total corneal refractive power in 3 and 4 mm zones (PTCRP3 and PTCRP4), and Aladdin keratometry (AK). Regression-based models mapping the measured C0 and C45 components (Alpin's method) to reconstructed corneal astigmatism were derived.
Results: Mean C0 components were 0.50/0.59/0.51 dioptres (D) for IOLMK/PK/AK; 0.2/0.26/0.31 D for IOLMTK/PTCRP3/PTCRP4; and 0.26 D for reconstructed corneal astigmatism. All corresponding C45 components ranged around 0. The prediction models had main diagonal elements lower than 1 with some crosstalk between C0 and C45 (nonzero off-diagonal elements). Root-mean-squared residuals were 0.44/0.45/0.48/0.51/0.50/0.47 D for IOLMK/IOLMTK/PK/PTCRP3/PTCRP4/AK.
Conclusions: Results from the different modalities are not consistent. On average IOLMTK/PTCRP3/PTCRP4 match reconstructed corneal astigmatism, whereas IOLMK/PK/AK show systematic C0 offsets of around 0.25 D. IOLMTK/PTCRP3/PTCRP4. Prediction models can reduce but not fully eliminate residual astigmatism after toric IOL implantation.
| Original language | English |
|---|---|
| Pages (from-to) | 627-638 |
| Number of pages | 12 |
| Journal | Clinical & Experimental Ophthalmology |
| Volume | 52 |
| Issue number | 6 |
| DOIs | |
| Publication status | Published - Aug 2024 |