Effect of Glucose Reference Measurement Distribution on MARD Performance Measure for CGM

Harald Kirchsteiger, Florian Reiterer, Guido Freckmann, Luigi Del Re

Research output: Contribution to journalArticlepeer-review

Abstract

The mean absolute relative difference (MARD) is commonly used toquantify the performance of CGM sensors. It is well known that there are significant differences in the MARD if different glucose ranges are considered (e.g. in hypo-, normo- or hyperglycemia). Here, we analyzed the effect of the distribution of glucose reference measurements on MARD.
Original languageEnglish
Pages (from-to)A38-A39
Number of pages2
JournalDiabetes Technology and Therapeutics
Volume17
Issue numberS1
Publication statusPublished - Feb 2015