Skip to main navigation Skip to search Skip to main content

A novel online recalibration strategy for continuous glucose measurement sensors employing LMI techniques

  • Harald Kirchsteiger (Speaker)

Activity: Talk or presentationContributed talkunknown

Description

This paper considers the problem of online calibration and recalibration of continuous glucose monitoring devices. A parametric relation between interstitial and blood glucose is investigated and a constructive algorithm to adaptively estimate the parameters within this relation is proposed. The algorithm explicitly considers measurement uncertainty of the device used to collect the calibration measurements and enables automatic detection of measurements which are not suitable to be used for calibration. The method was assessed on clinical data from 17 diabetic patients and the improvements with respect to the current state of the art is shown.
Period05 Jul 2013
Event titleEngineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Event typeConference
LocationJapanShow on map

Fields of science

  • 207109 Pollutant emission
  • 202027 Mechatronics
  • 206 Medical Engineering
  • 202003 Automation
  • 203 Mechanical Engineering
  • 202034 Control engineering
  • 202012 Electrical measurement technology
  • 203027 Internal combustion engines

JKU Focus areas

  • Mechatronics and Information Processing