Abstract
Continuous glucose monitoring (CGM) systems
are becoming increasingly popular for the management of type 1
diabetes mellitus (T1DM). However, one of the limitations of using
CGM information for blood glucose (BG) control is the fact
that the CGM sensor measures the glucose not in the blood, but
in the interstitial fluid. The current paper shows how a sensor
and patient-specific time delay between interstitial glucose (IG)
and blood glucose (BG) can be identified from CGM recordings
and BG measurements. The resulting time delay values were
found to be highly patient-specific and correlated with patient
characteristics. Furthermore, it was found that the CGM time
delays are a good predictor for the required total daily dose
(TDD) of insulin, as well as for the carbohydrate-to-insulin-ratio
(CIR). Based on these findings we introduce a method of how
insulin therapy in T1DM could possibly be adjusted based on
identified CGM time delays.
| Originalsprache | Englisch |
|---|---|
| Titel | IFMBE Proceedings, XIV Mediterranean Conference on Medical and Biological Engineering and Computing 2016 |
| Herausgeber*innen | Efthyvoulos Kyriacou, Stelios Christofides, Constantinos S. Pattichis |
| Verlag | Springer |
| Seiten | 190-195 |
| Seitenumfang | 6 |
| Band | 57 |
| ISBN (Print) | 978-3-319-32701-3 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - Apr. 2016 |
Publikationsreihe
| Name | IFMBE Proceedings |
|---|---|
| Band | 57 |
| ISSN (Print) | 1680-0737 |
UN SDGs
Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung
-
SDG 3 – Gute Gesundheit und Wohlergehen
Wissenschaftszweige
- 206002 Elektromedizinische Technik
- 207109 Schadstoffemission
- 202 Elektrotechnik, Elektronik, Informationstechnik
- 202027 Mechatronik
- 202034 Regelungstechnik
- 203027 Verbrennungskraftmaschinen
- 206001 Biomedizinische Technik
JKU-Schwerpunkte
- Mechatronics and Information Processing
Dieses zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver