Abstract
We outline calibrated measurements of the microwave reflection coefficient from the tunnel junction of an ultra-high vacuum low tem-
perature scanning tunneling microscope. The microwave circuit design is described in detail, including an interferometer for an enhanced
signal-to-noise ratio and a demodulation scheme for lock-in detection. A quantitative, in situ procedure for impedance calibration based on
the numerical three-error-term model is presented. Our procedure exploits the response of the microwave reflection signal due to the change
of the tunneling conductance caused by sub-nm variation of the tunneling distance. Experimental calibration is achieved by a least-squares
numerical fit of simultaneously measured conductance and microwave reflection retraction curves at finite conductance. Our method paves
the way for nanoscale microscopy and spectroscopy of dielectric surface properties at GHz frequencies and cryogenic temperatures. This
opens a promising pathway even for dielectric fingerprinting at the single molecule limit.
| Originalsprache | Englisch |
|---|---|
| Aufsatznummer | 103702 |
| Seiten (von - bis) | 103702/1-7 |
| Seitenumfang | 7 |
| Fachzeitschrift | Review of Scientific Instruments |
| Volume | 94 |
| Ausgabenummer | 10 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - 01 Okt. 2023 |
Wissenschaftszweige
- 210006 Nanotechnologie
- 103 Physik, Astronomie
- 103011 Halbleiterphysik
- 103018 Materialphysik
- 103009 Festkörperphysik
- 103017 Magnetismus
JKU-Schwerpunkte
- Sustainable Development: Responsible Technologies and Management
Dieses zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver